Recent Eppendorf Centrifuge 5403 Centrifuge, Table Top questions, problems & answers. Free expert DIY tips, support, troubleshooting help & repair advice for all Science & Laboratory. Learn more about Eppendorf® Refrigerated Microcentrifuge, Model 5417R. We enable science by offering product choice, services, process excellence and our people make it happen. Centrifuge incorporates reinforced steel rotor housing and double lid locks. The maintenance-free, brushless induction motor reaches maximum speed within 12 seconds. Eppendorf Centrifuge 5403 Manual Pdf-adds 1 ->>->>->> DOWNLOAD 1 / 3.
Advertising
Troubleshooting
Centrifuge 5418 / 5418 R English (EN)
32
7.2
Error messages
If one of the following error messages appears, proceed as follows:
1. Remove fault (see Remedies).
2. If necessary, repeat centrifugation.
Symptom/message Cause
Remedy
LID ERROR (5418)/
LID (5418 R)
Centrifuge lid cannot be locked.
Try again to close centrifuge lid.
LID ERROR (5418)/
LID (5418 R)
Centrifuge lid cannot be released.
1. Switch the centrifuge off and back on.
2. Press the
open key.
If the error occurs again:
1. Switch off centrifuge.
2. Activate the emergency lid release
Emergency lid release on p. 33).
LID ERROR (5418)/
LID (5418 R)
Centrifuge lid must not be released
during a run.
Wait for rotor to stop.
INT
Mains/power failure during a run.
Check the mains connection.
NO RPM (5418)/
Error
3 (5418 R)
Error in the speed measuring system or
drive overheated.
Leave the device switched on until the
error message disappears (10 s or 6 min).
Error 5
(only for 5418 R)
Prohibited opening of lid or lid switch is
defective during a run.
1. Wait for rotor to stop.
2. Open and close again the lid of the
device.
3. Repeat the run.
Err 6 (5418)/
Error
6 (5418 R)
Drive fault.
Repeat the run.
If this error message appears again,
switch centrifuge off and back on again after > 20 s.
Error 6
(only for 5418 R)
Drive overheated.
Allow the drive to cool down for at
least 15 min.
Err 7 (5418)/
Error
7 (5418 R)
Major deviation in the speed control.
1. Wait for rotor to stop.
2. Tighten rotor.
Err 8 (5418)/
Error
8 (5418 R)
Drive fault.
1. Wait for rotor to stop.
2. Repeat the run.
Err 9 to 17 (5418)/
Error
9 to
17 (5418
R)
Electronics error.
Eppendorf 5417r Centrifuge Manual
Switch centrifuge off and back on
again after > 20 s.
Error 18
(only for 5418 R)
Too high temperature deviation from set
value in the rotor chamber.
Eppendorf Centrifuge Manual
Check the settings.
Check unhindered air circulation
through the air slots.
Thaw ice or switch off device and
allow it to cool down.